Labconco TXE Tall XPerEnclosure 3892022 User Manual

Page 34

Advertising
background image

Chapter 6: Maintaining Your TXE

Product Service 1-800-522-7658

30

1. Push the ENTER button on the face of the alarm until the Set up Menu is

displayed.

2. Scroll to SET UP and hit ENTER.

3. The PASSWORD MENU displays (The Password is 0000). Press the

ENTER button repeatedly until the CAL CONFIG MENU is displayed.

4. In the CAL CONFIG MENU set for the following:

CALIBRATION CONFIG MENU

SETTINGS

DISPLAY UNITS

FPM

LOW AIR ALARM

SEE CHART ABOVE – LOW AIR ALARM

LOW AIR CUTOFF

OFF

LOW AIR CUTOFF

SEE CHART ABOVE – LOW AIR ALARM

WARNING AIR ALARM

SEE CHART ABOVE – LOW CALIBRATION SET
POINT (Set 3-5 FPM above low air alarm)

WARNING AIR RESET

3 FPM

HIGH AIR ALARM

OFF

HIGH AIR ALARM

OFF

LOWER AIR SAMPLE FLUCTUATIONS

3%

HIGHER AIR SAMPLE FLUCTUATIONS 3%
LOWER/HIGHER AIR SAMPLE
DIFFERENCE

10 FPM

WARN TO ALARM AIR TIME

10 SECONDS

ALARM TO WARN AIR TIME

3 SECONDS

SHOW AIR FLOW

ON

SHOW TIME LINE OFF = DISPLAYS
BAR GRAPH

OFF

AUDIBLE ALARM

ENABLED

SENSOR DIFFERENCE

2%

5. To complete the CAL CONFIG, be sure to enter “DONE”. You are

returned to the Main Menu.

Calibrate Alarm

1. Scroll to set up and hit ENTER, then CALIBRATE and hit ENTER.

2. The PASSWORD MENU displays (The Password is 0000). Press the

ENTER button repeatedly until the next menu is displayed.

3. Adjust the speed of the fan until the low airflow velocity is reached. Use

your thermal anemometer readings to determine the low air velocity.

4. Enter the value of your low airflow velocity using your +/- scroll keys. Be

sure the front panel is installed. Step out of the way, wait approximately 30
seconds for air currents to settle and hit ENTER. The airflow alarm will
sample the airflow.