Initial certification, Re-certification, Fluorescent light replacement – Labconco TXE Tall XPerEnclosure 3892022 User Manual

Page 36

Advertising
background image

Chapter 6: Maintaining Your TXE

Product Service 1-800-522-7658

32

Initial Certification

The TXE has been certified at the factory for an inflow velocity of 90±10 fpm
along with the HEPA Filter Leak Test. The TXE enclosure should be certified for
the proper inflow velocity required by your Safety Officer. It is also a
conservative recommendation to perform the HEPA Filter Leak Test again should
there be any damage caused during transport.

Re-Certification

Under normal operating conditions, the enclosure should be recertified at least
annually if serviced. The certifier should perform the following tests.

Inflow Velocity Test

HEPA Filter Leak Test


In addition, the following tests should also be performed at the user’s discretion:

Measure of Line Voltage and Current

Smoke Test to determine proper airflow patterns

Lighting Intensity Test (when appropriate)

Noise Level Test (when appropriate)

Vibration Test (when appropriate)

Fluorescent Light Replacement

1. Disconnect the power.

2. Locate the light reflector located under the control panel shown in Figure

6-2 and Figure 6-3.

3. Remove the light reflector support by removing four Phillips screws on the

side of the light reflector.

4. Rotate and remove the old fluorescent lamp.

5. Reinstall the new fluorescent lamp and light reflector in reverse order.

6. Power the unit up and try the new fluorescent lamp.

Advertising