Before you install the TXE, you need to prepare your site for installation. You
must be certain that the area is level and of solid construction. In addition, a
dedicated source of electrical power should be located near the installation site to
power the TXE, balance and other apparatus. Additionally, the enclosure should
be strategically placed in the lab to provide efficient workflow.

Carefully read this chapter to learn the requirements for your installation site:

The location and air current requirements.

The exhaust and blower requirements.

The electrical power requirements.

The space requirements.

Refer to Appendix B: Dimensions for complete enclosure dimensions.

Refer to Appendix C: Specifications for complete TXE electrical and
environmental conditions, specifications and requirements.

Location and Air Current Requirements

The TXE has been designed to contain hazards by negating typical cross drafts and
turbulence within the opening. As a precautionary safety measure and a higher
level of quality management, it is recommended that the enclosure be placed in an
area to avoid:

High traffic areas where walking might cause an air disturbance or be a
nuisance to balance readings.

Overhead or wall HVAC diffusers, fans, radiators or other lab equipment
producing air currents.

Next to doorways or windows that may be opened.

Do not position the unit so that it is difficult to operate the main
disconnect device.

Ne placez pas l'appareil de sorte qu'il est difficile de faire
fonctionner le dispositif principal de déconnexion.

!