Guardian digital 1000 airflow monitor, Programming the guardian 1000 digital alarm – Labconco TXE Tall XPerEnclosure 3892022 User Manual

Page 33

Advertising
background image

Chapter 6: Maintaining Your TXE

Product Service 1-800-522-7658

29

Guardian Digital 1000 Airflow Monitor

Option

Refer to Figure 6-3 for operation and calibration. Refer to the separate Labconco
1000 Alarm Manual for detailed information.

Operation

The Guardian Digital Airflow Monitor consists of the airflow sensor, the Alarm Unit,
and the 15 VDC power supply. For 115V operation the alarm unit is powered by
plugging the power supply into the factory-prepared digital airflow monitor socket.
For 230V operation, the Alarm Unit is powered by plugging the power supply into a
building outlet. The alarm has “Enter,” “+,” and “-” buttons to program the monitor.
There is also a green LED “SAFE,” yellow LED “CAUTION,” and red LED “LOW”
with audible alarm for airflow conditions. The audible alarm can be permanently
muted if desired. The Guardian Digital 1000 Airflow Monitor displays a face
velocity value, provides an RS232 communications port to a PC or building computer
system, can be configured for external input connections such as night setback or
external alarm and provides up to three output relays that can be configured. For
complete detailed information, please refer to the separate Labconco 1000 Alarm
User’s Manual provided with the enclosure.

Calibration

The TXE comes factory calibrated with the monitor alarm at 60 ± 10 fpm with the
inflow velocity set at 90± 10 fpm.

Programming the Guardian 1000 Digital Alarm

Set the desired face velocity of the enclosure with the doors closed using an
average face velocity of readings from a thermal anemometer.

Configure Alarm

To successfully calibrate the airflow monitor, it will be necessary to change the
face velocity by adjusting the airflow exhaust volume. The exhaust volume can be
adjusted using the speed control located behind the front panel on the TXE.

The following inflow face velocity speeds are recommended to successfully
calibrate. To begin, determine what the low airflow, low warning airflow, normal
airflow, and high airflow shall be.

Low Air Alarm

Set Point

(user defined)

Enclosure

Operating

Inflow Speed

Low Calibration

Set Point

(user defined)

High Calibration

Set Point

(user defined)

60 – 70 fpm

80 fpm

50 – 90 fpm

100 – 150 fpm

80 - 90 fpm

100 fpm

50 – 110 fpm

100 – 170 fpm


NOTE: ENTER button stores information

+/- Buttons allow for scrolling.

Advertising