Acids must not be used, Safety precautions – Labconco FilterMatPortable Exhausters 3970024 User Manual

Page 29

Advertising
background image

25

Chapter 4: Using Your FilterMate

Product Service 1-800-522-7658

The FilterMate Chemical Guide chart provides
estimated saturation times up to the TWA
concentration for the Chemical.

In addition to the above, all other guidelines for
chemical use in the FilterMate apply
.

Acids Must Not Be Used

The FilterMate blower cannot be exposed to acids. Where
applications require the use of acids, the enclosure must be vented
to the outside by a remote blower without the use of the
FilterMate.

Safety Precautions

NOTICE: The following safety precautions
must be strictly followed. Refer to Application
Section and Chemicals Appropriate for Use for
additional safety information

• Ensure that the FilterMate is connected to an electrical

service in accordance with local and national electrical
codes. Failure to do so may create a fire or electrical
hazard. Do not remove or service any electrical
components without first disconnecting the FilterMate
from electrical service.

• Use of an open flame must be avoided with the

FilterMate. Open flames may disrupt the airflow
patterns in the enclosure and cause a fire hazard with
volatile solvents.

• Increase the face velocity or change the HEPA filter

when the face velocity falls below acceptable limits
established by your Safety Officer.

• Exhausted carbon filters and HEPA filters are to be

disposed of as hazardous waste. The user is responsible
for recording the chemicals adsorbed or removed by the
filters and disposing properly. HEPA filters should be
disposed of with the disposal bag connected to the rear
of the FilterMate.

!

Advertising