Ou u r r, At t e e – Labconco FilterMatPortable Exhausters 3970024 User Manual

Page 41

Advertising
background image

37

Product Service 1-800-522-7658




C

C

H

H

A

A

P

P

T

T

E

E

R

R

6

6

M

M

O

O

D

D

I

I

F

F

Y

Y

I

I

N

N

G

G

Y

Y

O

O

U

U

R

R

F

F

I

I

L

L

T

T

E

E

R

R

M

M

A

A

T

T

E

E


The following procedure may be necessary to modify the
FilterMate.

Converting a Room Exhaust
HEPA FilterMate to Outside
Exhaust


Refer to Figures 3-2 and 3-3.

1. Remove the perforated exhaust cover and replace it with

the FilterMate 6" canopy outside exhaust cover (Catalog
#3913600).

2. Next connect the ductwork to the remote blower. Remote

blowers available from Labconco are listed below:

3. Remote Blowers (Roof-Mounted) and Dampers. 1/4 hp

direct drive. Corrosion-resistant epoxy-coated steel
housing and wheel. Blower inlet is 6.00" ID. Outlet
dimensions are 4.25" x 7.38" OD. See dimensional data in
enclosure manual.

CFM @ Static Pressure – Inches of H

2

O

S.P. 0.0" 0.125" 0.25" 0.50" 0.75" 0.87"
CFM

595 560 515 420 300 167

Catalog #

Description

Shipping
Wt. lbs./kg

4863500

Remote Blower, 115 volts, 60 Hz, 4.4 amps

35/16

4863501

Remote Blower, 115/230 volts, 50 Hz, 5.6/2.8 amps

35/16

7053501

Explosion-Proof Remote Blower, 115 volts, 60 Hz,
4.4 amps

40/18

3924000

6" Powder Coated Metal In-line adjustable damper

4724200

6" PVC In-Line adjustable damper

Advertising