6
Chapter 2: Prerequisites
Product Service 1-800-522-7658
CAUTION: This equipment must be grounded
(earthed).
Ground Fault Interrupter
WHEN CONNECTING THE FILTERMATE TO A
GROUND FAULT INTERRUPTER CIRCUIT
(GFIC), ENSURE THAT THE CIRCUIT IS
RATED FOR 15-AMP LOAD AND IS
FUNCTIONING CORRECTLY. CONNECTION
OF THE FILTERING SYSTEM TO A LOWER
CAPACITY OR DEFECTIVE GFIC MAY
RESULT IN EXCESSIVE ACTIVATION OF THE
GFIC, CUTTING POWER TO THE
FILTERMATE WHILE IT IS OPERATING.
Exhaust Requirements for the
FilterMate (Filtered Room
Exhaust or Outside)
The FilterMate is capable of exhausting up to 280 cfm of HEPA
filtered air or up to 220 cfm of combination HEPA/Carbon filtered
air. These airflows are achieved when connected to the enclosure
with an 8' length of 5" hose. For lower airflows adjust the speed
control on the back of the FilterMate. Labconco offers HEPA and
Carbon FilterMate models listed below.
FilterMate
Catalog
Number Voltage
Filter
Exhaust
Airflow
(cfm) (max)
Exhaust
Connection
3970000
115 Volt/60 Hz
HEPA
280
None
3970001
115 Volt/60 Hz
Carbon
280
None
3970002
115 Volt/60 Hz
HEPA
280
Canopy to outside
3970003
115 Volt/60 Hz
HEPA/Carbon
220
None
3970004
115 Volt/60 Hz
Carbon/Carbon
220
None
3970020
230 Volt/50 Hz
HEPA
280
None
3970021
230 Volt/50 Hz
Carbon
280
None
3970022
230 Volt/50 Hz
HEPA
280
Canopy to outside
3970023
230 Volt/50 Hz
HEPA/Carbon
220
None
3970024
230 Volt/50 Hz
Carbon/Carbon
220
None
!
)