Ground fault interrupter – Labconco FilterMatPortable Exhausters 3970024 User Manual

Page 10

Advertising
background image

6

Chapter 2: Prerequisites

Product Service 1-800-522-7658

CAUTION: This equipment must be grounded
(earthed).

Ground Fault Interrupter

WHEN CONNECTING THE FILTERMATE TO A
GROUND FAULT INTERRUPTER CIRCUIT
(GFIC), ENSURE THAT THE CIRCUIT IS
RATED FOR 15-AMP LOAD AND IS
FUNCTIONING CORRECTLY. CONNECTION
OF THE FILTERING SYSTEM TO A LOWER
CAPACITY OR DEFECTIVE GFIC MAY
RESULT IN EXCESSIVE ACTIVATION OF THE
GFIC, CUTTING POWER TO THE
FILTERMATE WHILE IT IS OPERATING.

Exhaust Requirements for the
FilterMate (Filtered Room
Exhaust or Outside)

The FilterMate is capable of exhausting up to 280 cfm of HEPA
filtered air or up to 220 cfm of combination HEPA/Carbon filtered
air. These airflows are achieved when connected to the enclosure
with an 8' length of 5" hose. For lower airflows adjust the speed
control on the back of the FilterMate. Labconco offers HEPA and
Carbon FilterMate models listed below.

FilterMate

Catalog
Number Voltage

Filter

Exhaust
Airflow
(cfm) (max)

Exhaust
Connection

3970000

115 Volt/60 Hz

HEPA

280

None

3970001

115 Volt/60 Hz

Carbon

280

None

3970002

115 Volt/60 Hz

HEPA

280

Canopy to outside

3970003

115 Volt/60 Hz

HEPA/Carbon

220

None

3970004

115 Volt/60 Hz

Carbon/Carbon

220

None

3970020

230 Volt/50 Hz

HEPA

280

None

3970021

230 Volt/50 Hz

Carbon

280

None

3970022

230 Volt/50 Hz

HEPA

280

Canopy to outside

3970023

230 Volt/50 Hz

HEPA/Carbon

220

None

3970024

230 Volt/50 Hz

Carbon/Carbon

220

None

!

)