Refer to the following if your FilterMate fails to operate properly.
If the suggested corrective actions do not solve your problem,
contact Labconco for additional assistance.

PROBLEM CAUSE CORRECTIVE

ACTION

Blower won’t turn
on

Unit not plugged into outlet

Plug the FilterMate into appropriate
electrical service

Circuit breaker(s) or Ground
Fault Interrupter

Reset circuit breaker

Blower wiring is disconnected Inspect blower wiring

Blower switch is defective

Replace switch

Motorized impeller is defective Replace motorized impeller

Low face velocity
or poor
containment of
contaminants

Enclosure sash not closed.

Close sash to the lowest position

HEPA Filter clogged

Replace HEPA filter or increase
speed

Carbon filter loaded with
chemicals

Replace carbon filter