This Appendix contains technical information about the FilterMate
including specifications, environmental operating conditions,
wiring diagrams, dimensions, and evaporation rates.
Electrical Specifications

• FilterMate 10 Amps, 115V, 50/60 Hz.

• FilterMate, 10 Amps, 230V, 50/60 Hz.


Environmental Conditions

• Indoor use only.

• Maximum altitude: 6562 feet (2000 meters).

• Ambient temperature range: 41° to 104° (5° to 40°C).

• Maximum relative humidity: 80% for temperatures up to 88°F

(31°C), decreasing linearly to 50% relative humidity at 104°F
(40°C).

• Main supply voltage fluctuations not to exceed ±10% of the

nominal voltage.

• Transient overvoltages according to Installation Categories II

(overvoltage categories per IEC 1010). Temporary voltage
spikes on the AC input line that may be as high as 1500V for
115V models and 2500V for 230V models are allowed.

• Used in an environment of Pollution degrees 2 (i.e., where

normally only non-conductive atmospheres are present).
Occasionally, however, a temporary conductivity caused by
condensation must be expected, in accordance with IEC 664.