Chapter 5: Using The Ductless Enclosure

Product Service 1-800-522-7658

32

Diagnostic Program Screens





























Pressure Sensor Diagnostic Notes:

X20 CAL example represents a

pressure of 134 with the impeller off

and 212 with the motor speed at 157

producing 90 fpm for this calibration

example. With Smart Flow enabled,

the Paramount maintains constant

pressure as indicated by the pitot

tube and pressure sensor to maintain

constant airflow face velocity.

Organic Sensor Diagnostic Notes:

Organic Sensor example represents

sensor S1 calibrated in clean air at

1400 millivolts. When sensor S1

sees an increase of 400 millivolts to

1800 millivolts, the Early Warning

“Replace Filter-s” appears on the

LCD display with an intermittent beep

as only 5-10% filter life remains.

Filters should be changed upon

completion of work. When sensor S1

sees an increase of 500 millivolts to

1900 millivolts, the Final “Replace

Filter-s” Alarm sounds a continuous

beep that can only be muted for 5

minutes.