This Appendix contains technical information about all the Paramount
Ductless Enclosures including electrical specifications and environmental
operating conditions.

Electrical Specifications

All 2', 3', 4', 5' Paramounts – 100-115V, 5 Amps, 50/60 Hz or 230V,

3 Amps, 50/60 Hz.

6' Paramounts – 100-115V, 10 Amps, 50/60 Hz or 230V, 5 Amps,

50/60 Hz.

Environmental Conditions

Indoor use only.

Maximum altitude: 10,000 feet (3,058 meters). Airflow must be

recalibrated at altitudes above or below 900 feet (275 meters).

Ambient temperature range: 41° to 104°F (5° to 40°C).

Maximum relative humidity: 80% for temperatures up to 88°F

(31°C), decreasing linearly to 50% relative humidity at 104°F (40°C).

Main supply voltage fluctuations not to exceed ±10% of the nominal

voltage.

Transient overvoltages according to Installation Categories II

(Overvoltage Categories per IEC 1010). Temporary voltage spikes on
the AC input line that may be as high as 1500V for 115V models and
2500V for 230V models are allowed.

Used in an environment of Pollution degrees 2 (i.e., where normally

only non-conductive atmospheres are present). Occasionally,
however, a temporary conductivity caused by condensation must be
expected, in accordance with IEC 664.