Chapter 6: Maintaining the Ductless Enclosure

Product Service 1-800-522-7658

49

Organic Sensor Board Replacement

If the organic sensor board is replaced, the Paramount must be calibrated for
the “Organic Sensor.” Contact Labconco Product Service. See Appendix A:
Replacement Parts
and reference Figure 6-5. Please note that the 6'
Paramount utilizes two organic sensor boards, one for each motorized
impeller.

1.

Consult the wiring diagram on the back of the front panel and disconnect

all power.

2.

Remove the organic sensor connector as shown in Figure 6-5.

3.

Remove the four Phillips screws that mount the organic sensor board.

4.

Install the new organic sensor board in reverse order. Note: Be careful

with the delicate electronic sensors on the organic sensor board and do
not move them or rotate them.

5.

Follow the “Calibration” menu screens from Chapter 5 as the Organic

Sensor needs calibration. A separate code and procedure is required and
Product Service must be contacted.

Figure 6-5

Organic Sensor Board