Chapter 5: Using The Ductless Enclosure

Product Service 1-800-522-7658

37

Work Techniques

Keep all materials a minimum of 4 inches (100 mm) from the inside

of the lower air foil, and perform all contaminated operations as far to
the rear of the work area as possible.

Segregate all clean and contaminated materials in the work area.

Arrange materials to minimize the movement of contaminated

materials into clean areas.

Keep all discarded contaminated material to the rear of the work area.

Avoid moving materials or the operator's hands and arms through the

front access opening during use.

Avoid the use of an open flame.

Avoid using techniques or procedures that disrupt the airflow patterns

of the enclosure.

Final Purging

Upon completion of work, the enclosure should be allowed to operate

for two to three minutes undisturbed, to purge airborne contaminants
from the work area.

Unloading Materials and Equipment

Objects in contact with contaminated material should be surface

decontaminated before removal from the enclosure.

All open trays or containers should be covered before being removed

from the enclosure.

Wipe-Down

Wipe down the interior surfaces of the enclosure with suitable

solvent, disinfectant, or cleaner. Important Note: Clean up
procedures using alcohols or volatile chemicals with low filter
capacity could saturate the filters quickly.

Dispose of used rubber gloves appropriately, and have

soiled/contaminated lab coats laundered properly.

Wash hands and arms thoroughly with soap.

Shutdown

Turn off the fluorescent light and enclosure blower to preserve filter life.