Labconco Paramoun Ductless Enclosures 69636 Series, 6' Widths User Manual

Page 40

Advertising
background image

Chapter 5: Using The Ductless Enclosure

Product Service 1-800-522-7658

35

Filter Setup Display Options

This menu selection determines which filters are loaded and displayed in
the lower and upper positions. One of seven filter types must be selected
for each filter position.

The filter check time and filter final life must be selected next for each
filter position. Consult Chapter 3, Appendix E, the Labconco Specialist,
and the Paramount Chemical Guide for proper program times.

Restoring Software Default Settings

To restore the default settings to their original configuration:

1.

Access the Menu, and select “SETUP.” Press “ENTER.”

2.

In the Setup submenu, select “Restore Defaults.” Press “ENTER.”

Table of Default Settings

Blower:

Smart-Flow enabled

Units of Measure:

fpm

Security Lock:

OFF, Non-secure

Timers:

Blower and Light Timers disabled.

Filter Timers:

Enabled; Interval set at 100 hours, Final set at

500 hours.

Calibration

This password-protected selection is for qualified certifiers to adjust the
blower speed, face velocity, and the low airflow alarm setpoint. Certifier
access code is set at 123.

Diagnostic Test Operation

When the “Diagnostic Test” menu is selected, the LCD allows “Keypad
Test” to help troubleshoot any malfunction. The “Pressure Sensor”
displays pressure values to help troubleshoot an airflow malfunction.
The “Organic Sensor” displays chemical breakthrough (millivolt
readings) and alarm setpoints to troubleshoot an organic sensor
malfunction.