Refer to the following table if your ductless enclosure fails to operate
properly. If the suggested corrective actions do not solve your problem,
contact Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

“Replace Filters”
continuous alarm
occurs

Improper final time
programmed and
display reads “Replace
Filters-t”, (t = time)

See Chapters 3 and 5 to properly set the
Filter Final Life Timer.

Organic sensor alarms
and display reads
“Replace Filters-s”,
(s = organic sensor)

Organic sensor has detected filter
breakthrough. Replace filters
immediately. A spare set of filters
should always be on hand. Consult
Chapters 3, 4, and 5, the Labconco
Specialist, and the Chemical Guide for
estimated filter life based on specific
chemical use. All chemicals have a
different estimated filter life based on
specific chemical use. Always limit
chemical use, cover up chemicals when
not in use and turn off blower to
preserve filter life.

“Mute” will not stay
operational

The “Mute” alarm can only be off for 5
minutes and must be pushed every 5
minutes until the filters have been
replaced. See Chapters 3, 5 & 6.

“Check Filters”
intermittent alarm
occurs frequently

Improper check time
programmed and
display reads “Check
Filters-t”, (t = time)

See Chapters 3 and 5 to properly set the
Filter Check Life Timer.