Chapter 6: Maintaining the Ductless Enclosure

Product Service 1-800-522-7658

48

Main Control Board Replacement

If the main control board is replaced, the Paramount must be calibrated per
the menu screens in Chapter 5. See Appendix A: Replacement Parts, and
reference Figure 6-4.

1.

Consult the wiring diagram on the back of the front panel and disconnect

all power.

2.

Remove the hardware to rotate the header control panel.

3.

Locate the connectors to the main control board and remove the

connectors per the wiring diagram.

4.

Label the two pressure hoses and remove the two pressure hoses from the

control board. The hoses must be installed properly with the front hose
corresponding with the front “total pressure” hose from the motorized
impeller assembly (front to front throughout).

5.

Remove the control board from the plastic display bezel by removing the

four Phillips screws.

6.

Reassemble the new control board in reverse order and follow the

“Calibration” menu screens from Chapter 5 as the Pressure Sensor,
Airflow Alarm, and Organic Sensor all need calibration.