Chapter 6: Maintaining Your Filtered PCR Enclosure

Product Service 1-800-522-7658

31

5. Adjust the downflow velocity to the nominal operating point required. The

green “safe” LED should light.

6. Over time the HEPA filter will load and eventually slow the downflow

velocity. Once the alarm condition is met, simply increase the speed
control as outlined in Chapter 6 or replace the HEPA filter if the speed
control is maximized.

7. The table below lists typical alarm conditions based on normal operating

conditions. Typical alarm conditions are set at downflow velocities of 10
to 30 feet per minute below the normal operating conditions due to supply
air and exhaust air fluctuations, as well as room air cross drafts. Consult
your Safety Officer or Laboratory Manager for proper operating speeds.

Enclosure Operating

Downflow Speed

Alarm Condition Set Point

Speed

55 ± 10 fpm

25-35 fpm

45 ± 10 fpm

25-35 fpm