Chapter 6: Maintaining Your Filtered PCR Enclosure

Product Service 1-800-522-7658

33

Initial Certification

The Purifier Filtered PCR Enclosure has been certified at the factory for an
average downflow velocity of 40-55 fpm along with the HEPA Filter Leak Test.
The Purifier Filtered PCR Enclosures should be certified for the proper downflow
velocity required by your Safety Officer. It is also a conservative recommendation
to perform the HEPA Filter Leak Test again should there be any damage caused
during transport.

Re-Certification

Under normal operating conditions, the Purifier Filtered PCR Enclosure should be
recertified at least annually. The certifier should perform the minimum following
tests:

Downflow Velocity Test

HEPA Filter Leak Test

In addition, the following tests should also be performed at the user’s discretion:

Measure of Line Voltage and Current

Smoke Test to determine proper airflow patterns

Lighting Intensity Test (when appropriate)

Noise Level Test (when appropriate)

Vibration Test (when appropriate)

Fluorescent Light Replacement

1. Disconnect the power.
2. Locate the small light reflector located under the control panel.
3. Remove the light reflector by removing two Phillips screws on the bottom

of the light reflector.

4. Rotate and remove the old fluorescent lamp.
5. Reinstall the new fluorescent lamp and light reflector in reverse order.
6. Power the unit up and try the new fluorescent lamp.

UV Light Replacement

1. Disconnect the power.
2. Locate the UV lamp located inside the enclosure.
3. Rotate and remove the old UV lamp.
4. Reinstall the new UV lamp.
5. Power the unit up and try the new UV lamp (UV closure panel must be

installed).