Ppendix, Iltered, Nclosure – Labconco Purifier Filtered PCR Enclosures 3970425 User Manual

Page 54: Pecifications

Advertising
background image

Product Service 1-800-522-7658

50





A

PPENDIX

C

F

ILTERED

PCR

E

NCLOSURE

S

PECIFICATIONS

This Appendix contains technical information about all Purifier Filtered PCR
Enclosures including electrical specifications and environmental operating conditions.

Electrical Specifications

5 Amps, 115V, 60 Hz

3 Amps, 230V, 50/60 Hz

Environmental Conditions

Indoor use only.

Maximum altitude: 6562 feet (2000 meters).

Ambient temperature range: 41° to 104°F (5° to 40°C).

Maximum relative humidity: 80% for temperatures up to 88°F (31°C),

decreasing linearly to 50% relative humidity at 104°F (40°C).

Main supply voltage fluctuations not to exceed ±10% of the nominal

voltage.

Transient over-voltages according to Installation Categories II (Over-

voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.

Used in an environment of Pollution degrees 2 (i.e., where normally only

non-conductive atmospheres are present). Occasionally, however, a temporary
conductivity caused by condensation must be expected, in accordance with
IEC 664.