Chapter 3: Getting Started


After you verify the fume hood components, move your hood to
the location where you want to install it. Then, follow the steps
listed next to remove the shipping skid from your unit.


1. Remove the side panels by unscrewing the Phillips

screws.

2. Find the hardware (bolts, washers, nuts) that attach the

fume hood to the skid and remove the hardware. Some
hardware is on the sides and some is on the back.

Sash Weight Release

To protect the fume hood from damage in shipment, the sash
weight has been secured to the back of the fume hood with four (4)
screws. Simply remove the screws and make sure the sash cables
are on the pulleys before operating the sash. On models with more
than one sash, the sash weights have been secured to the shipping
skid with lag screws. Remove the weights from the skid and attach
them to the respective sash cables using the hooks provided.

NOTE: THE SASH WEIGHT ITSELF WAS
INDIVIDUALLY MATCHED FOR THIS
SPECIFIC HOOD AND SHOULD NOT BE
EXCHANGED ON ANY OTHER UNIT.

Install the Protector XStream
Hood on a Supporting Structure
and Work Surface

The Protector XStream Hood is heavy! Use
caution when lifting or moving the unit.


When installing the Protector XStream Fume Hood onto a
chemically-resistant work surface or benchtop, ensure that the
structure can safely support the combined weight of the fume hood
and any related equipment. The work surface should be at least as
wide as the hood to properly support it.

The work surface is aligned flush with the back
of the fume hood for good airflow: this will
provide the correct spacing under the air foil for
proper bypass airflow. The Protector XStream
performs best with a work surface having a large

!

!

)