Labconco Protector XStream Laboratory Fume Hoods 9840803 User Manual

Page 29

Advertising
background image

25

Product Service 1-800-522-7658

Chapter 4: High Performance Features and Safety


Always work as far back in hood as possible.
It is best to keep all chemicals and apparatus
6" inside the front of the hood. This hood
does not feature explosion-proof electrical
components, unless ordered separately.
Therefore, use of flammable or explosive
materials in quantities above the explosive
limit are not recommended.

Do not work with chemicals in this hood
without the exhaust system running. Do not
store chemicals in a fume hood.

Perchloric acid use in this hood is prohibited.

High level radioisotope materials are
prohibited for usage in this hood.

AVOID CROSS DRAFTS AND LIMIT
TRAFFIC IN FRONT OF THE HOOD. AIR
DISTURBANCES CREATED MAY DRAW
FUMES OUT OF THE HOOD.

The use of heat-generating equipment in this
hood without the exhaust system operating
properly can cause damage to the hood.

The Protector XStream Laboratory
Hood should be certified by a qualified
certification technician before it is
initially used. The hood should be re-
certified whenever it is relocated,
serviced or at least annually
thereafter.


Ensure that the unit is connected to electrical
service in accordance with local and national
electrical codes. Failure to do so may create
a fire or electrical hazard. Do not remove or
service any electrical components without
first disconnecting the hood from electrical
service.

)

!

!

Advertising