Routine service operations – Labconco Protector XStream Laboratory Fume Hoods 9840803 User Manual

Page 36

Advertising
background image

32

Product Service 1-800-522-7658

Chapter 6: Maintaining Your Protector Fume Hood


Monthly (or more often as required)

• Determine the actual face velocity through the sash opening

of the hood where the average reading should be at the
specified velocity. (Use calibrated thermal anemometer or
other approved apparatus).

• Using a damp cloth, clean the exterior surfaces of the hood,

particularly the front of the hood, to remove any
accumulated dust.

• Check all service valves, if so equipped, for proper

operation.

• The hood baffles should be checked for blockages behind

them to ensure that the hood is maintaining proper airflow.

• All weekly activities.


Annually

• Replace the fluorescent lamps.

• Have the fume hood recertified by a qualified certification

technician. See Certifying the Protector XStream Fume
Hood in Chapter 3
.

• All monthly activities.


Biannually

• The sash assembly should be checked to ensure proper

operation and to make sure there are no signs of abnormal
wear on the sash pulleys, cables and clamps.

Routine Service Operations


Front Panel Removal:

1. Simply lift the front panel up and then away from the hood to

provide access to the top.


Changing the Fluorescent Lamp:

1. Turn light switch to “OFF”.
2. Remove the front panel as noted earlier.
3. Reach over the front header of the hood and remove the knock

out plugs at both ends of fixture. Lift fixture up.

4. Remove the fluorescent lamp by pushing it out of the spring-

loaded lamp socket and swinging it out of the other lamp
socket.

5. Install the new lamp by reversing the removal procedure.

S

A