Operation checklist, Vacuum pump ballast setting, Operating the freeze dryer – Labconco FreeZone 4.5 Liter Freeze Dry Systems 77510 Series User Manual

Page 20: Set-up

Advertising
background image

Chapter 4: Using Your Freeze Dryer

Product Service 1-800-522-7658

16

11.

Collector Temperature Graph Display – This display indicates the temperature of the

collector. The highest LED indicates the collector temperature is warmer than 10°C.
The indicators will sequence down when the temperature reaches 10, 0, -10, -20, -30, -
40°C. When the collector temperature is –40°C or lower the green indicator will light.

12.

Alarm Indicator – This red LED indicates that a system alarm has occurred. Press the

Menu Switch to display the alarm message on the LCD display.

13.

Main Power Switch – Turns the Freeze Dryer on or off. (Not shown, located on the

right side of the cabinet.)

Operation Checklist

The following checklist should be followed prior to each use of your Freeze Dryer:

1.

Wipe the interior of the collector chamber with a soft cloth or paper towel to

remove any accumulated moisture.

2.

Check the collector chamber drain hose to ensure that the hose is free of

moisture and that the drain plug is securely installed. The Freeze Dryer will
not start if moisture is detected.

3.

Using a soft, lint-free cloth or paper towel, wipe the collector chamber lid

gasket to remove any dirt and contaminants that could cause a vacuum leak.
Vacuum grease is not required on the lid gasket to obtain a proper vacuum seal.

4.

Check that each sample valve is closed or in the “vent” position.

Vacuum Pump Ballast Setting

Most vacuum pumps are equipped with a gas ballast mechanism. The freeze dry
process requires high vacuum. Therefore, it is recommended that the gas ballast be
closed during the operation of the Freeze Dry System. If the gas ballast is left
open for extended periods of operation, the oil can be pumped out the exhaust,
causing the pump to fail.

Operating the Freeze Dryer

Set-Up


The Freeze Dryer may be configured to automatically start the vacuum pump when
the collector temperature reaches –40°C. The display units for vacuum may be
selected to be mBar, Pascal (Pa) or Torr and the temperature may be displayed as
°F or °C. The run time of the refrigeration system and the vacuum pump may be
monitored.

To configure your Freeze Dryer, turn the main power switch ON and press MENU.
The display will show:





VACUUM UNITS

mBar

Pa

Torr

Advertising