Chapter 8: Modifying Your Freeze Dryer Display

Product Service 1-800-522-7658

35

Moisture Sensor Alarm

The moisture sensor alarm may be disabled. It is shipped enabled as the factory
default setting.

Modifying the Display

To enter an offset or to configure the alarms, perform the following steps:

Press and hold MENU button (approximately 10 seconds) until the display
shows:



It is recommended that you do NOT perform a Master Reset because all
offsets and configurations that have been entered will be lost. If you must
perform a Master Reset, hold SELECT approximately 5 seconds.

This will reset:

Vacuum units to mBar
Vacuum set point to 0 mBar
Temperature units to °C
Refrigeration total hours to 0
Vacuum pump total hours to 0
RS232 transmission rate to 10 seconds
Temperature offset to 0°C
Vacuum offset to 0 mBar
Low line voltage to 0 V offset
High line voltage to 0 V offset
Moisture sensor to enabled
Refrigeration to operate in Manual Mode


The display will default to the main freeze dry display information.

If you do not want to restore all factory settings, but want to adjust offsets, press
MENU briefly and the display will show:



Press SELECT until the appropriate offset is shown. Each number
represents 1°C

PRODUCTION MENU
MASTER RESET?

COLLECTOR TEMP OFFSET:
O C