TROUBLESHOOTING

TROUBLESHOOTING TIPS
The following table provides troubleshooting tips for using the AED Trainer 3 If you need additional assistance, please
contact your local laerdal representative or go to www.laerdal.com/aedt3

Issue

Possible Cause

Suggested Action

AED Trainer 3 does not work
and gives beeping signals when
turned on.

• No Operating System

Card is installed.

• Install the Operating System Card.

• Incorrect card is

installed.

• Install the original Operating System Card.

• Operating System Card

has corrupt or missing
files.

• Obtain replacement Operating System

Card.

AED Trainer 3 gives
“Analyzing” prompt before
pads are applied to manikin.

• The HeartStart Training

Pads II are connected;
the AED Trainer 3
senses that pads are
in place as soon as the
pads connector plug is
inserted.

• To use HeartStart Training Pads II, apply

the pads to the manikin before connecting
them to the AED Trainer 3. Alternatively,
use HeartStart Training Pads III or the
Laerdal LINK system, so that the AED
Trainer 3 automatically detects when pads
are correctly placed.

AED Trainer 3 does not
respond when pads are
applied to manikin.

• Laerdal LINK

Technology Training
Pads are incorrectly
placed on the Laerdal
LINK Technology
manikin.

• Correct the position of the pads.

AED Trainer 3 battery
depletes prematurely.

• One or more of the

batteries is incorrectly
oriented.

Reinstall batteries as illustrated inside the
battery compartment.