Setup, Cleaning the system – Laerdal SimMan Essential User Manual

Page 18

Advertising
background image

18

SETUP

Cleaning the System

After each session and before storage, clean the IV Arm system:

1 Make sure the collector bag is connected to the drain tube.

2 Connect a syringe filled with Isopropanol 60%-70% to the IV

Catheter. Press the prime switch, and flush the IV Arm system
with Isopropanol. Release the prime switch when done.

3 Connect a syringe filled with air to the IV Catheter. Press the

prime switch, and flush the IV Arm system until only air exits the
drain tube. Release the prime switch when done.