Troubleshooting, Clinical features - pneumothorax bladders, Simulator shut down – Laerdal SimMan Essential User Manual

Page 37: Troubleshooting the iv arm

Advertising
background image

37

Clinical Features - Pneumothorax Bladders

Problem

– Experience problems with Pneumothorax Bladders.

Solution

– Check tubing connection at base of the bladder, to ensure that

tubing has not been disconnected.

Simulator Shut down

Problem

– The Patient Simulator is unresponsive.

Solution

Press and hold the <ON/OFF> button for 10 seconds to force the
Patient Simulator to shut down.

Troubleshooting the IV Arm

Problem

– Veins do not energize.

Solution:

– Ensure Patient Simulator is powered on and IV Arm cable and air

tube are connected to the Patient Simulator.

– Ensure the blood reservoir is filled, see Filling and priming the

system.

– Ensure the tourniquet area is activated (depressed).

– Look for a leak between the IV Arm pad and the IV Arm chassis.

– If the tourniquet area has been activated several times within a

short period of time, internal pressure in the IV Arm may drop.
Wait 30 seconds to let pressure rebuild and try again.

Problem

– When injecting more than 40ml fluid and prime switch is

depressed, no fluid is visible in the drain tube.

Solution

– Leak inside IV Arm. Contact Laerdal Technical Service.

Problem

– After releasing the prime switch and filling the internal reservoir,

more than 16ml of fluid can be injected into the arm.

Solution

– Leak inside the IV Arm. Contact Laerdal Technical service

Problem

– There is a leak between IV Arm Pad and IV Arm chassis.

Solution

– Ensure the IV Arm Pad is fully inserted and secure in the IV Arm

chassis.

– If this does not help, too high pressure has been applied to the

IV Arm Pad during the filling process. Replace the IV Arm Pad.

Problem

– No flashback when inserting Intravenous Catheter into IV Arm

Pad

Solution

– Ensure that the collection bag is at the same elevation as the IV

Arm.

TROUBLESHOOTING