-- Supply network analyzers CVM-BD-420-4 & CVM-BD-420-8 -- M98132301-20 --- Page Nº 26


5.10.- 4 - 20 mA outputs: CVM-BD-420-4

(4 analog outputs) and

CVM-BD-420-8

(8 analog outputs)

With these outputs we can configure the CVM-BD420 to give an output of 4 - 20

mA d.c. or of 0 - 20 mA

d.c. ( resolution of 4.000 points ) proportional to any of

the parameters measured by the CVM-BD, with the ability of setting the scale
(offset and full scale).

On the CVM-BD-420-.. screen following messages appear at this SET-UP point
(provided the right module is connected to the equipment):

a.- Parameter choice:

dA 1

OUTPUT D/A No.1

Code

xx

Parameter No.


-

"max" -- "min" keys: allow the selection of any parameter from 1 to 30, the 35, or
from 54 to 59.

- “display” key: validates the selected option and passes to the next setup screen.


b.- Selection of 0 - 20 mA or 4 - 20 mA :

dA 1

OUTPUT D/A Nr.1

Scal

Scale :

4 - 20

allows choosing a 0 - 20 mA

or 4 - 20 output ("max" or "min" key)


- "display": to validate the selected option and pass to the next setup screen.