AMETEK i-iX Series II User Manual
Page 94

User Manual
California Instruments
94
i Series II / iX Series II / iM Series II
VOLT SENSE
This field selects the internal or external sense line inputs.
Internal sense does not require the external sense lines to be
connected as sensing occurs at the output relay of the AC
source. For best results, connect the external sense lines and
select the EXT sense mode in this field. This will compensate
for voltage drop in the cables to the load. The measurements
are also taken at the sense points, so the external sense mode
should be used for best measurement results.
NO. OUTPUT
This field is only accessible if the phase mode option is
installed. It allows the number of output phases to be toggled
between single and three-phase mode. In single-phase mode,
all power is available on phase A. The Mode-iX option is
available to route all power to a single output terminal pair.