SCPI Command Reference
Understanding SCPI Commands

100

Operating Manual for Multichannel Functionality (GPIB-M)

The following punctuation is sent with the command string:

•

Colons (:) separate command keywords from lower-level keywords.
For example, CAL:CURR:STAT.

•

Blank spaces separate command keywords from parameter values.
For example, CURR 0.1.

•

Commas separate parameters from each other when more than one parameter is
sent in the same string. For example, CAL:STAT OFF,”1234.”

•

Semicolons (;) separate multiple commands from the same subsystem. This
allows for greater efficiency. For example:

CAL:CURR:LEV MIN;VOLT:LEV MIN

is the same as typing:

CAL:CURR:LEV MIN

CAL:VOLT:LEV MIN

•

Colons and semicolons can be used together to link commands from different
subsystems. For example:

CAL:CURR:LEV MIN;:MEAS:CURR?

The following punctuation is not sent with the command string:

•

Braces ( { } ) identify a selection of choices. Choose one of the enclosed values.

•

Vertical bars, or pipes, ( | ) separate the choices found within the braces.

•

Angle brackets ( < > ) identify where specific values must be entered for a
parameter. For example, in the example at the top of the page, the parameter
<current> appears in the command string. To set the current setpoint to 0.1A, the
syntax is CAL:CURR:LEV 0.1.

•

Square brackets ( [ ] ) identify optional parameters or keywords. If an optional
parameter is not sent with the command string, a default parameter will be used.

Using Minimum and Maximum

In the following example, Minimum and Maximum are offered as alternative choices
to declaring a specific parameter value.

CAL:CURRent:LEVel {<current>|MIN|MAX}

The string CAL:CURR:LEV MIN sets the current calibration level to the minimum
model value.