Sorensen SLD-Series DC Load

Operation

M540075-01 Rev C

3-3

4 REM

LED

Indicates remote operation:

•

When lit, the unit is under remote control and cannot be operated through the front
panel keys.

•

When off, the unit is under local control and can be manually operated using the
front panel keys.

5 NG A LED

When enabled and Channel A is selected (CHAN A LED is lit), indicates whether or not
Voltage and/or Current exceed(s) the upper or lower limit set for channel A.

•

Lit = “No Go” – Channel A Voltage and/or Current limits exceeded.

•

Not lit = "Go" (no fail) for Channel A.

NOTE:

Use the CONFIG key, 14, to set NG ON/OFF to ON, to enable this function.

6

NG B LED

When enabled and channel B is selected (CHAN B LED is lit), indicates whether or not
Voltage and/or Current exceed(s) the upper or lower limit set for channel B.

•

Lit = “No Go” – Channel B Voltage and/or Current limits exceeded.

•

Not lit = "Go" (no fail) for Channel B.

NOTE:

Use the CONFIG key, 14, to set NG ON/OFF to ON, to enable this function.

7

Upper 5-Character Display and Associated LEDs

Multi-purpose display, depending on selected mode:

•

In PRESet OFF mode, displays voltage of DC input terminal or V

SENSE

BNC input.

•

In PRESet ON mode, displays values being programmed for CR or CV.

•

In DYN STTING On mode, displays dynamic waveform parameters for:

◦ High Level Current (Curr-H)
◦ Low Level Current (Curr-L)
◦ High Load Duration (T-Hi)
◦ Low Load Duration (T-Lo)

◦ Slew Rate (Sr)

•

In LIMIT ON mode:

◦ displays GO/NG limit values for voltage

◦ displays “AH” or “AL” when current limits a being programmed (see DYN

STTING key, 12)

•

In CONFIG ON mode:

◦ displays SENSE to indicate V

SENSE

control

◦ displays voltage settings for LOAD ON or for LOAD OFF