2 over current, 3 over power, 4 over temperature – AMETEK SLH Series RevC User Manual

Page 52

Advertising
background image

Operation

Sorensen SLH-Series DC High Power Electronic Load

3.10.2 Over Current

•

The protection point of over current protection (OCP) is preset in the load. The
activation values of the OCP circuit are shown in Table 3-3.

•

When an over-current condition occurs, the OCP circuit is activated, and the lower 5-
character monitor on the front panel of the electronic load flashes “oCP”. The load will
automatically go to a LOAD OFF state.

•

Once the over current condition is reset, the lower 5-character monitor will resume
normal operation.

3.10.3 Over Power

•

The protection point of over power protection (OPP) is preset in the load. The activation
values of the OPP circuit are shown in Table 3-3.

•

When an over-power condition occurs, the OPP circuit is activated, and the upper 5-
character monitor on the front panel of the electronic load flashes “oPP”. The load will
automatically go to a LOAD OFF state.

•

Once the over power condition is reset, the upper 5-character monitor will resume
normal operation.

3.10.4 Over Temperature

•

SLH-Series DC electronic loads also are designed with a temperature sensor(s) for over
temperature protection (OTP). The activation values are shown in Table 3-3.

•

When an over-temperature condition occurs, the OTP circuit is activated, and the lower
5-character monitor on the front panel of the electronic load flashes “otP”. The load will
automatically go to a LOAD OFF state.

•

When OTP occurs, please check the environment condition such as the ambient
temperature and distance between the rear panel of Electronic load mainframe and wall
is greater than 6in. / 15cm.

•

Once the over temperature condition is reset, the lower 5-character monitor will resume
normal operation.

Model OVP

OCP

OPP

OTP

SLH-60--120-600 63V 126A 630W 85

°C

SLH-60-120-1200 63V 126A 1260W 85

°C

SLH-60-120-1800 63V 126A 1890W 85

°C

SLH-60-240-1200 63V 252A 1260W 85

°C

SLH-60-240-1800 63V 252A 1890W 85

°C

SLH-60-360-1800 63V 378A 1890W 85

°C

Table 3-3 Summary of Over Protection Values

3-22