Sorensen SLH-Series DC High Power Electronic Load
Remote Programming
bit 7
bit 0
7
6
5
4
3
2
1
0
Limited
Range Changed
Invalid Command
Invalid Operating
Figure 5-3 Error Status Byte Register
bit 0:Limited
This bit set to high by Electronic Module Load setting command causing over range. Reset
by CLER command.
Example:
SLH-60-30-150
MODE CC
RANG 1
CURR:HIGH 20.0
setting current=20.0 A
actual setting current=3.071 A
bit 0 will set to high
bit 1:Range Changed
This bit set to high by Electronic Module Load setting command causing change range.
Reset by CLER command.
Example:
SLH-60-30-150
FALL 0.050 MA
FALL 100.0 MA
setting fall time to 100.0 MA will change range then bit 1 will set to high.
bit 2 :invalid command
This bit set to high by accepted illegal command.
bit 3 :invalid operating
This bit set to high by accepted invalid operating command.
Example:
SLH-60-120-1200
MODE CR
DYN ON
DYNAMIC function only supports CC MODE.
VOLTAGE METER
Purpose:
The reading of 4 1/2 digit voltage meter read back query command.
Command Syntax:
M540074-01
Rev C
5-17