Cleaning and storing the unit, Error messages/trouble-shooting, Battery – Beurer BM 85 User Manual

Page 32

Advertising
background image

32

If your smartphone has a protective cover, remove this to ensu-

re that there is no interference during the transfer. Begin the

data transfer in the Beurer “HealthManager” app.

9. Cleaning and storing the unit

• Clean the unit and cuff carefully using a slightly damp cloth

only.

• Do not use any cleaning agents or solvents.

• Under no circumstances should you hold the unit under wa-

ter, as this can cause liquid to enter and damage the unit.

• If you store the unit, no heavy objects should be placed on

top of it. The cuff line should not be bent sharply.

10. Error messages/trouble-shooting

In case of faults, the _ message appears in the display.

Error messages may appear if:

• systolic or diastolic pressure could not be measured (

or

appears on the display)

• systolic or diastolic pressure was outside the measurement

range ( or

Lo appears on the display)

• the cuff is fastened too tightly or loosely (

or

appears

on the display)

• the pump pressure is higher than 300 mmHg (

appears

on the display)

• pumping up takes longer than 160 seconds (

appears on

the display)

• there is a system or device error (

,

,

or

ap-

pears on the display)

• the data could not be sent to the PC (

appears in the

display).

• the data could not be sent via Bluetooth

®

( ).

In such cases, repeat the measurement and/or the data trans-

fer. Ensure that the cuff line is correctly attached and that you

do not move or speak during the measurement.

Technical alarm – description

Should the recorded blood pressure (systolic or diastolic) lie

outside the limits specified in the section “Technical specifica-

tions”, the technical alarm will appear on the display indicating

either “

” or “ Lo ”. In such cases, you should seek medical as-

sistance and check the accuracy of your procedure.

The limit values for the technical alarm are factory set and can-

not be adjusted or deactivated. These alarm limit values are

accorded second priority under the standard IEC 60601-1-8.

The technical alarm is a non-locking alarm and must not be

reset. The signal shown on the display will disappear automati-

cally after about 8 seconds.

11. Battery

The BM 85 is equipped with a lithium ion battery (3.7V/400

mAh). If the

symbol flashes, you must charge the battery

for a minimum of two hours using the enclosed cable. Rechar-

ge the battery to between 50 and 75 % of its capacity at least

twice a year to achieve a maximum battery service life.

Device charged (75% - 100%)

Device charged (50% – 75%)

Device charged (25% - 50%)