12. Technical specifications

Model no.

BM 85

Measurement me-

thod

Oscillometric, non-invasive blood pres-

sure measurement on the upper arm

Measurement range

Cuff pressure 0 – 300 mmHg,

systolic 60 – 260 mmHg,

diastolic 40 – 199 mmHg,

pulse 40 –180 beats /minute

Display accuracy

Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement inac-

curacy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /

diastolic 8 mmHg

Memory

2 x 60 memory spaces

Dimensions

L 180 mm x W 99 mm x H 40 mm

Weight

Approximately 317 g (without cuff)

Cuff size

22 to 36 cm

Permissible operating

conditions

+ 10 °C to + 40 °C, ≤ 90 % relative humi-

dity (non-condensing)

Permissible storage

conditions

-20 °C to + 55 °C, ≤ 90 % relative humi-

dity, 800 –1050 hPa ambient pressure

Power supply

DC 5 V

600mA lithium ion battery

3.7 V/400mAh

Battery life

For approx. 50 measurements, depen-

ding on levels of blood pressure and

pump pressure

Accessories

Cuff, cuff holder, instructions for use,

USB mains part, USB cable, storage

pouch

Classification

Internal supply, IPX0, no AP or APG,

continuous operation, application part

type BF

Data transfer via

Bluetooth® wireless

technology

The blood pressure monitor uses Blue-

tooth® Smart (low energy),

2.4 GHz frequency band,

Compatible with Bluetooth 4.0 smart-

phones/tablets

List of supported smartphones/tablets