Note

The blood glucose monitor is made of precision components. The accuracy of the measurements

and service life of the device depend on its careful handling:

• Protect the device from impacts and do not drop it.

• Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or

water, marked temperature fluctuations, direct sunlight and extreme cold.

• Do not use the device in the vicinity of strong electromagnetic fields, radio systems or mobile

telephones.

8 w

hat

if

there

are

problems

?

Display messages on batteries and blood glucose measurement

No.

Cause

Solution

LP

Batteries empty.

Replace all batteries.

Ht

Temperature of the measuring envi-

ronment, blood glucose monitor or

test strip above the permitted range.

Repeat the test using a new test strip as soon

as the measuring environment, blood glucose

monitor and test strips have reached room tem-

perature (+20°C to +26°C).

Lt

Temperature of the measuring envi-

ronment, blood glucose monitor or

test strip below the permitted range.

Repeat the test using a new test strip as soon

as the measuring environment, blood glucose

monitor and test strips have reached room tem-

perature (+20°C to +26°C).

Err

Used or contaminated test strip in-

serted.

• Insert an unused test strip that has not ex-

pired.

• Repeat the blood glucose measurement.

Err 001

System error.

Remove batteries, reinsert batteries. Contact

Customer Services if the problem persists.

Err 002

Insufficient blood on the test strip.

Repeat the measurement using a new test

strip.

Err 005

System error.

Remove batteries, reinsert batteries. Contact

Customer Services if the problem persists.

Unknown error messages.

Remove batteries, reinsert batteries. Contact

Customer Services if the problem persists.