2.11 Supply pressure

Do not apply pressure until electrical connections are made. When applying pressure to the system, take

care to avoid pressure shocks in the system, and increase pressure gradually. Be sure that all setpoints are

at zero when doing so.

2.12 System purging

If corrosive or reactive fluids are used, purging with an inert gas is absolutely necessary, because if the tubing

has been exposed to air, introducing these fluids will tend to clog up or corrode the system due to a chemical

reaction with air or moist air.

Complete, long term purging is also required to remove such fluids from the system before exposing it to air.

It is preferred not to expose the system to air, when working with these corrosive fluids. Cycled purging and

evacuating is preferred.

2.13 Seals

Bronkhorst HIGH-TECH B.V. has gathered a material compatibility chart from a number of sources believed

to be reliable.

However, it is a general guide only. Operating conditions may substantially change the accuracy of this guide.

Therefore is no liability for damages accruing from the use of this guide. The customer’s application demands

its own specific design or test evaluation for optimum reliability. So check if the seals are correct for your

process. However the CEM system has metal seals for all connections. Only the valve plunger is an

elastomer.