B) to start the process, Procedure: actions, C) to stop the process – Bronkhorst CEM W-300B User Manual

Page 20: D) how to leave system overnight: procedure, Actions, E) to empty and purge the system

Advertising
background image

BRONKHORST HIGH-TECH B.V.

B) To start the process:

Procedure:

Actions:

B1)Set the CEM heater at right temperature.

Give setpoint to the heater, wait until

temperature is reached.

B2)

Start gas flow.

(1)

Open 9. Give setpoint to gas MFC.

B3) Start liquid flow.

(2)

Give setpoint to liquid flow controller, wait

until stabilization.

B4) Connect process.

Open 10, close 11.

C) To stop the process:

Procedure:

Actions:

C1) Disconnect process.

Open 11, close 10.

C2) Stop liquid flow.

Give setpoint = 0 to liquid flow controller.

C3) Optional: Stop gas flow.

(1)

Give setpoint = 0 to gas MFC.

D) How to leave system overnight:

Procedure:

Actions:

D1) Relief the supply pressure from the liquid

supply and lines.

Close 2, close 5 and open 1 for some

seconds.

Relief pressure inert gas supply, open 2.

D2) Leave system unused.

Close 3. Preferably, keep pumping the lines,

close 2, open 1.

D3) Before startup, pressurize liquid supply.

Close 1 and open 2.

Open 3 and 5.

D4) System is now ready for use.

Proceed according to B.

E) To empty and purge the system:

(3,4)

Procedure:

Actions:

E1) Follow procedure “How to leave system

overnight” until “leave system unused”.

See D1 and D2.

E2) Purge liquid out of lines.

Open 4 and 7.

E3) Purge lines with inert gas.

Close 1.

Open 2 briefly for some time.

E4) Evacuate liquid lines (cont.).

Close 2, open 1.

E5) Optional: Repeat last two steps for a few

times.


1

Make sure that the carrier gas is always flowing when the liquid flow is on. Preferably,

let always gas flow in the carrier gas line when the system is filled with liquid.

2

Close supply pressure of the liquid vessel during the process. The pressure controller of

the inert gassupply can influence the stability of the liquid flow controller.

3

If the liquid is reactive or aggressive, be sure to remove air (and water vapor) carefully, before

introducing liquid in the system.

4

If the liquid is reactive, aggressive, poisonous, toxic, or flammable, beware that the system is

purged fully (preferably with inert liquid, E3, E4, E5) before opening the system.

Remarks:

- In case of non-pure liquids, place a filter just before the Liqui-Flow.

- Mount the CEM system vertical.

App. 1, page 4