Startup without using vacuum, A) to fill the system, Procedure: actions – Bronkhorst CEM W-300B User Manual

Page 21

Advertising
background image

Startup without using Vacuum

Liquid

4

3

5

6

8

10

11

9

MFC

Carrier
Processgas

Vacuum

7

LFM

Inertgassupply

2

Process

Vacuum

Valve

Heatedline

CEM

Liquid

Gas

A) To fill the system:

Procedure:

Actions:

A1)Purge all lines

Close: 3, 5 and 10.

Connect inert gas on carrier gas inlet.

Open : 2, 4, 7, 9 and 11.

Give setpoint to gas controller, the control

valve 8 opens.

Give setpoint = 0 to liquid - flow controller

Purge for a long time.

A2) Pressurize liquid supply.

Close 4, give right pressure on inert gas

supply. Open 3.

A3) Fill the system with liquid.

(1)

Let gas flow in the carrier gas line during

filling of the system.

Open 5 briefly, allow mixing valve 6 to fill.

Wait until all gas bubbles left the system,

then close 7.

A4) The system is now ready for use.

1

Make sure that the carrier gas is always flowing when the liquid flow is on. Preferably,

let always gas flow in the carrier gas line when the system is filled with liquid.

App. 1, page 5