Flowserve 520MD Digital Positioner User Manual

Page 14

Advertising
background image

14

®

User Instructions Logix 520MD - LGENIM0520-01 09/09

Blink

Code

Description

Recommendations

YYYG

SUPPLY PRESSURE LOW WARning (user set) indicates that the

supply pressure is below the user set warning limit. Low supply

pressure can cause poor valve response or positioner failure.

The minimum recommended supply pressure is 30 PSI (2.1 bar)

for proper operation. The unit will fail at less than approximately

17 PSI (1.2 bars). Low supply pressure indications can also be

caused by pneumatic leak.

Regulate the supply pressure at the positioner above 30 PSI (2.1 bar). Reca-

librate pressure sensors. Ensure system air/gas supply is adequate. Repair

kinked supply tubing. Check the pressure sensor board connections and replace

pressure sensor board if necessary. Check for pneumatic leaks in the actuator

and actuator tubing. See DTM screen: Health Status/Actuator Health. This

indicator can be disabled.

YYYY

ACTUATiOn RATiO WARning (user set) indicates a decreased

ability of the system to actuate the valve. It is based on the ratio

of available force to required force to actuate. It is affected by the

process load, friction, spring force, and available supply pressure.

Increase the supply pressure. Reduce the friction. Check the actuator spring.

Resize the actuator. Adjust user set limits. See DTM screen: Health Status/

Actuator Health. This indicator can be disabled.

YRGG

PiLOT RELAY RESPOnSE WARning (user set) indicates that the

pilot relay is sticking or slow to respond. This affects the respon-

siveness, increases the chance of limit cycling and excessive air

consumption. The pilot relay is part of the inner loop and con-

sists of the driver module assembly with piezo (I-P relay) which

is coupled to the poppet. The value of this indicator corresponds

with inner loop lag. Delayed response can be caused by a partial-

ly clogged piezo or debris, oil, corrosion or low supply pressure.

Check response of the valve. If OK, adjust Pilot Relay Response limits. Check

supply pressure. Replace the piezo or driver module assembly. Maintain a clean,

water-free air/gas supply. See DTM screen: Health Status/Positioner Health.

This indicator can be disabled.

YRGY

FRiCTiOn LOW WARning (user set) indicates the friction has

passed below the user set limit.

Low friction is usually an indication of improperly loaded packing or seals in the

valve and actuator. See DTM screen: Health Status/Valve Health. This indicator can

be disabled.

YRGR

PnEUMATiC LEAK WARning (user set) indicates that the positi-

oner has detected a leak in the actuation assembly. Leakage from

the actuator can cause decreased responsiveness and excessive

air/gas consumption. Low supply pressure can also trigger this

warning.

Repair pneumatic leaks at the tubing junctions and actuator seals. Ensure proper

supply pressure. See DTM screen: Health Status/Actuator Health. This indicator

can be disabled.

YRYG

FRiCTiOn HigH WARning

(user set) indicates the valve/actuator

friction has passed the user set limit. High friction can cause

loop oscillations, poor position control, jerky motion, or valve

sticking. It can be caused by build-up from the process on the

stem, trim or seat , by a failing bearing or guides in the valve

and actuator, galling of the trim or stem, excessively tightened

packing, linkages, or other valve/actuator mechanical issues.

Determine if the friction is significantly interfering with the valve control. If not,

consider increasing the friction warning limit. Consider the following to reduce

friction: Stroke the valve to clear off build-up. Clear any external mechanical ob-

struction, loosen the packing, clean the stem, repair or replace the actuator. Highly

localized friction or very jerky travel can indicate internal galling. Repair or replace

internal valve components. See DTM screen: Health Status/Valve Health. This

indicator can be disabled.

YRRY

ELECTROniC inAbiLiTY TO FAiL SAFE WARning indicates that

the piezo may be damaged. This may prevent the proper failure

position upon loss of signal/power. This condition may occur

briefly on an air-to-close valve that is held for long periods of

time in the closed position, or and air-to-open valve held in the

open position.

If alarm persists for more than 30 minutes, the Piezo assembly is damaged and

should be replaced. This indicator can be disabled.

YRRR

PnEUMATiC inAbiLiTY TO FAiL SAFE WARning indicates that

upon loss of air supply, the valve may not move to the fail-safe

position. The spring alone is not adequate to overcome the fric-

tion and process load in the system. The failsafe spring may have

failed, or it was not sized properly for the application. Friction or

process load may have increased.

Check for high friction. Repair or replace actuator spring. Reduce process load.

This indicator can be disabled.

RGGY

FEEDbACK READing PRObLEM DURing CALibRATiOn ALARM

indicates that during calibration, the range of motion of the positi-

on feedback arm was too small for optimum performance, or the

position sensor was out of range.

Check for loose linkages and/or adjust the feedback pin to a position closer to

the follower arm pivot to create a larger angle of rotation if the feedback rotation

is less than 15 degrees for the total valve travel and recalibrate. Briefly pushing

the quick-cal button acknowledges this condition and the positioner will operate

using the current short stroke calibration if otherwise a good calibration. If the

condition does not clear then adjust the positioner mounting, linkage or feedback

potentiometer to move the position sensor back into range then restart the cali-

bration. This error may be cleared by briefly pushing the quick-cal button, which

will force the positioner to use the parameters from the last good calibration.

RGGR

innER LOOP OFFSET TiME OUT ALARM

during calibration the

Inner Loop Offset value did not settle. This could result in less

accurate positioning.

Repeat the stroke calibration to get a more accurate ILO value. To proceed using

the less accurate ILO value, this error may be cleared by briefly pushing the

quick-cal button. Lowering the gain setting may help if the actuator is unstable

during the calibration. Gain settings can be physically adjusted on the device. A

lower letter represents lower gain.

RGYG

nOn-SETTLE TiME OUT ALARM

indicates that during calibration,

the position feedback sensor did not settle.

Check for loose linkages or a loose positioner sensor. This error may be cleared

by briefly pushing the quick-cal button, which will force the positioner to use

the parameters from the last good calibration. This error may appear on some

very small actuators during the initial calibration. Recalibrating may clear the

problem.

Advertising