Flowserve 520MD Digital Positioner User Manual

Page 9

Advertising
background image

9

®

User Instructions Logix 520MD - LGENIM0520-01 09/09

Indicator arrow

3. Pos. Characterization
Select Linear if the actuator position should be directly

proportional to the input signal.

Select Optional if another characteristic is desired, which

is set in conjunction with the next switch, labeled Optional
Pos. Char
.

4. Optional Pos. Char. – If the Pos. Characterization switch

is set to optional, this switch is active with the following

options:

The =% option will characterize the actuator

response to the input signal based on a standard 30:1 equal

percent rangability curve.

If Custom is selected, the positioner will be characterized to

a custom table that must be set-up ValveSight DTM software

or a HART handheld.

5. Auto Tune – This switch controls whether the positioner will

auto tune itself every time the Quick-Cal button is pressed

On enables an auto tune feature that will automatically de-

termine the positioner gain settings every time a Quick-Cal

is performed based on the setting of the rotary Gain switch.

nOTE: there is a small black arrow indicating the selection.

The slot does not indicate the chosen gain.

If the rotary Gain selector switch is set to E with

the Auto Tune switch On, a Flowserve nominal

response tuning set will be calculated and used.

If the rotary Gain selector switch is set to D, C, B, or A with

the Auto Tune switch On, progressively lower gain settings

will be calculated and used.

If the rotary Gain selector switch is set to F, G, or H with

the Auto Tune switch On, progressively higher gain settings

will be calculated and used.

Off forces the positioner to use one of the factory preset

tuning sets determined by the rotary Gain selector switch.

Settings A through H are progressively higher gain prede-

fined tuning sets. The positioner is preset to Autotune - On

at the factory. This is the recomended setting.

The gain selector operates as a “live” switch. This means

that changes to the switch position while the positioner is in

normal operation will have immediate effect on the control

algorithm.

6. Stability Switch – This switch adjusts the position control

algorithm of the positioner for use with low friction control

valves or high friction automated valves

Placing the switch to the left optimizes the response

for low friction, high performance control valves. This

setting provides for optimum response times when used

with most low friction control valves.

Placing the switch to the right optimizes the response

for valves and actuators with high friction levels. This

setting slightly slows the response and will normally

stop limit cycling that can occur on high friction valves.

9.3 Calibration switches
9.3.1 Setup of the Cal DIP Switch for the Quick
Calibration operating mode
Select Auto if the valve/actuator assembly has an internal

stop in the 100% stroke position. In Auto mode the

positioner will fully close the valve and register the 0%

position and then open the valve to the stop to register

the 100% position when performing a self-calibration.

See detailed instructions in the next section on how to

perform an auto positioner calibration.

Select Jog if the valve/actuator assembly has no cali-

bration stop in the open position. In the Jog mode the

positioner will fully close the valve for the 0% position

and then wait for the user to set the open position using

the Jog buttons labeled with the up and down arrows.

See the detailed instructions in the next section on how

to perform a manual calibration using the Jog buttons.

WARning: During the Quick-Cal operation the valve

may stroke unexpectedly. Notify proper personnel

that the valve will stroke, and make sure the valve is

properly isolated.

9.4 Quick-Cal Operation
The Quick-Cal button is used to locally initiate a

calibration of the positioner. Pressing and holding the

Quick-Cal button for approximately three seconds will

initiate the calibration. If the Config-Switches option is

enabled, the settings of all the configuration switches

are read and the operation of the positioner adjusted

accordingly. The Gain Selector switch is also read and

action will be taken to adjust the gain according to the

settings of the calibration switches as described in the

previous section. A Quick-Cal can be aborted at any

time by briefly pressing the Quick-Cal button and the

previous settings will be retained.

If the Quick calibration switch (not to be confused

with the Quick-Cal button) is set to Auto and the valve/

actuator assembly has the necessary internal stops, the

calibration will complete automatically. While the cali-

bration is in progress you will notice a series of different

lights flashing indicating the calibration progress. When

the lights return to a sequence that starts with a green

light, the calibration is complete. (See the appendix for

an explanation of the various light sequences.)

STOP!

Advertising