Flowserve APEX A9000 Accord User Manual

Page 22

Advertising
background image

22





Lose prev value? OK?

A warning that the previously set value will be lost.

Confirm with OK.

The calibration sequence starts

In service? Press OK

Calibration finished. Press OK to start positioner functioning. (If ESC is pressed, the positioner
assumes the

”Out of service” position but the calibration is retained).

Perform

Setting gain

Normal A,B,C,D,E,F,G See Page 17 for details on

how to set gain.






ExpertCal

Set point LO: Use the calibrator set to 4 Ma

(or set another value on the display). Press OK.

Set point HI: Use a calibrator of 20 mA

(or set another value on the display). Press OK.

Transmitter: Connect 10 - 28VDC. Connect an external mA meter to the loop. Read
low value on mA meter and adjust with up/down key. Press OK to set low value.
Repeat procedure

to set High value. Also see video on www.flowserve.com

Pot: Potentiometer setting, see section 8.

Also see video on www.flowserve.com

Full reset: Resets all set values.

Optional

Pressure LO: Use a supply of 2 bar (30 psi)

(or set another value on the display). Press OK. Pressure read

out only possible on Apex A9000

with built in pressure sensor.

Pressure HI: Use a supply of 7 bar (105 psi)

(or set another value on the display). Press OK. Pressure read

out only possible on Apex A9000

with built in pressure sensor.


















Advertising
This manual is related to the following products: