Flowserve LY Series Limitorque User Manual

Page 20

Advertising
background image

Limitorque LY Series: LY 1001, LY 2001 and LY 3001 FCD LMENIM1501-00 – 11/11

20

Figure 5.7 – Limit Switch Setting Nut

4. Rotate the CLOSE Limit Cam CW through the high side to the trip point until the Switch Plunger is

released. This causes the N.O. (normally open) contact to open.

NOTE:

There will be no electrical continuity at the trip point when measuring with an ohmmeter

between the common lead and the N.O. lead. You may also hear a faint “click” at the trip point.

NOTE:

During actuator operation, when the Switch Plunger trips, the N.O. contact is released,

causing the Limit Switch to stop the actuator in the CLOSED position.

Figure 5.8 – Setting CLOSE Limit Cam

5. Retighten Setting Nut (piece #7-10).

PERCENT

OPEN

0

50 100

Limitorque

7-10

Standard OPEN

Microswitch location

Standard CLOSE

Microswitch location

CLOSE microswitch

position while valve

is traveling from OPEN

toward the CLOSED

position

High Side

Rotate CW to

reach CLOSE

Trip Point

CLOSE

Limit Cam

CLOSE microswitch

position when valve

has reached the CLOSED

position and microswitch

has tripped

High Side

CLOSE

Trip Point

position

(“click”)

CLOSE

Limit Cam

Switch

Plunger

Switch Plunger

NO

NC

COM

NO

NC

COM

Advertising