Connect drains – Fluke Biomedical IDA-1S User Manual

Page 18

Advertising
background image

IDA-1S

Users Manual

8

• When you connect to the inlet circuits (for example, when you attach the

priming syringes to the stopcocks) make sure no new bubbles are
introduced.

Caution

To prevent possible damage to the product or to equipment
under test, do not use delivery set or components that have
been used for prior testing for patient infusion.

Note

Before you use the delivery set (the tubing, the syringe, and
stopcocks), make sure it is within the specified use period of the
manufacturer. Many sets are made to be used only once.

Connect Drains

Figure 4 shows a drain tube connected to the Product.

hqf04.eps

Figure 4. Drain Connections to the Product

Follow these recommendations when you connect the drain tubes to the Product
outlets:

• Do not allow the drain tubes to rise more than 10 cm (4 in) at any point above

the height of the inlet ports of the Product.

• The discharge end of the drain tubes must not be more than 10 cm (4 in)

below the bottom of the Product.