View Settings

Use View Settings to monitor the operation of the optical detectors on the
measuring burette. The Sensor column lists the sensor numbers from the top to
the bottom of the burette. The ADC column lists the output of each sensor. The
values change to reflect the water position in the burette. Use this information to
help diagnose fluid detection issues. Touch Utilities to return to the Utilities
menu.

Set Clock

Use Set Clock to set the date and time for your location. Touch the field and then
touch + or - to change the value.

Report Heading

Use Report Heading to enter a maximum of three lines of text. This text shows
when saved records are transferred to a PC. Fluke Biomedical recommends
using the first line for the name of the establishment. Touch a line and use the
on-screen keyboard to enter up to 20 characters per line. Touch Utilities to
return to the Utilities menu.

Instructions

The Instructions selection displays brief operating instructions for the Product.
Touch the Next and Prev (Previous) buttons to step through the instructions.

Calibration History

Calibration History retrieves the calibration change history records from the
measuring module and displays the details. Use the arrow buttons to scroll
through the list. Touch Utilities to return to the Utilities menu.

Touch Calibration

Use Touch Calibration to do a touch calibration process. Use a stylus to touch
the reference points on the screen. Touch Utilities to cancel and return to the
Utilities menu. The screen layout and calibration points follow the
recommendations of the touch panel manufacturers.