Product operation, Preferences, How to set up for a test – Fluke Biomedical IDA-1S User Manual

Page 19

Advertising
background image

Infusion Device Analyzer

Product Operation

9

Product Operation

The Product has rechargeable batteries that can supply a minimum of eight
hours of operation. The Product can operate while it charges. Control the Product
with the LCD touch screen.

Before you turn on the Product, make sure the Product calibration is up-to-date,
check for signs of wear, and ensure the battery door is closed.

To turn on the product and display the Start menu:

1. Push

the

WAKE UP button. If the startup screen appears without errors, the

Product is ready to use.

2. Touch

PRESS TO PROCEED.

The Start menu has these options:

Flow
Occlusion
Utilities
Shut Down

Preferences

Before you use the Product for the first time, set your preferences for display
functions, date and time, and the name of your location. The instructions for how
to set preferences are in the Utilities section in this manual.

How to Set Up for a Test

Each test starts with an information screen that contains data for the test. You
can change the data on the information screen for each test.

To change the data on the information screen:

1. Touch a field and use the on-screen keyboard to enter the data.

2. Use the buttons on the information screen to move between screens, to save

data, and delete data.