Occlusion tests – Fluke Biomedical IDA-1S User Manual

Page 23

Advertising
background image

Infusion Device Analyzer

Product Operation

13

Occlusion Tests

To do an occlusion test:

1. Select

Occlusion from the Start menu.

2. Complete the fields on the Occlusion Information screen. See the How to Set

Up for a Test section in this manual.

3. Touch

Next and follow the on-screen instructions. See Table 6 for an

example of the Occlusion Measurement screen.

4. When the test is complete, touch End. The Product prompts you to save the

results.

Note

You must remove the tube from the occlusion inlet, before you touch
ZERO to make sure that the Product senses the pressure correctly.

The initial display shows the pressure field is 0 (zero). The units of pressure
display as selected on the User Preferences screen. See the Utilities section
in this manual.

Table 6. Occlusion Measurement Screen

7.05

INST PRESSURE

(mmHg)

OCCLUSION

END

8.05

PEAK PRESSURE

(mmHg)

TIME OF PEAK

00:05:49

ELAPSED TIME

00:06:52

2

1

hqf08.eps

Item Description

The current pressure detected by the Product.

The peak pressure and the time the pressure was detected.

End

Stops the test. The Product prompts you to save the test.

Advertising