Flow tests – Fluke Biomedical IDA-1S User Manual

Page 21

Advertising
background image

Infusion Device Analyzer

Product Operation

11

Flow Tests

To do a flow test:

1. Select

Flow from the Start menu.

2. Complete the fields on the Flow Information screen. See the How to Set Up

for a Test section in this manual.

3. Touch

Next. The Flow Measurement screen appears. See Table 5.

Table 5. Flow Measurement Screen

0.00

DEVIATION %

0.00

AVE FLOW RATE (ml /hr)

0.00

INST FLOW RATE (ml/hr)

FLOW

PRIME

END

0.00

VOLUME (ml)

ELAPSED TIME

00:00:00

1

2

3

4

hqf06.eps

Item Description

The average and instantaneous (current) flow rates.

The deviation of the average flow rate from the set flow rate.

The prime indicator. When the Product senses fluid in the measuring tube, the prime
indicator is blue. If the Product does not sense fluid, the prime indicator is red.

The delivered volume and the elapsed time.

4. Prime the Product by continuously pushing water into the Flow inlet until the

prime indicator (the vertical bar) is all blue. Continue to push 5 ml into the
Flow inlet to make sure that no bubbles follow.

Caution

To avoid inaccurate readings, always repeat a test when
“Bubble” or “Air Lock” is shown on the display while a test is
running. See the Troubleshooting
section of this manual.