Infusion Device Analyzer

General Specifications

21

General Specifications

Battery Power ........................................................ 4 x Panasonic HHR210AB NiMh 2000 mAh batteries

Charger

Operating Voltage Range ................................... 100 V ac to 240 V ac

Supply Frequency ............................................... 50 Hz / 60 Hz

Supply Power ...................................................... <20 VA

Size (HxWxD) ......................................................... 30 cm x 17 cm x 10 cm (12 in x 8 in x 4 in)

Weight .................................................................... ~1.2 kg (2.7 lb)

Temperature

Operating ............................................................ 15

°C to 30 °C (59 °F to 86 °F)

Storage ............................................................... -20

°C to +40 °C (-4 °F to +104 °F) when drained of all liquid.

Humidity ................................................................. 10 % to 90 % non-condensing

Altitude ................................................................... 0 meters to 2000 meters (6500 feet)

Safety ..................................................................... IEC 61010-1: Overvoltage category II, Pollution Degree 2

Electromagnetic Environment ............................. IEC 61326-1: Basic

Emissions Classification...................................... IEC CISPR 11: Group 1, Class A. (Group 1 have intentionally

generated and/or use conductively coupled radio-frequency energy
which is necessary for the internal functioning of the equipment itself.
Class A equipment is suitable for use in non-domestic locations and/or
directly connected to a low-voltage power supply network.)

FCC ......................................................................... CFR47: Class A Part 15 subpart B

Storage of Results ................................................ Test results stored for later viewing, printing or transfer to PC. Typical

practical capacity: 100 tests

Power Down .......................................................... The results of tests in progress will be saved in the case of accidental

power down

Computer Control ................................................. The Product can be fully controlled from a PC using HydroGraph

software

Performance Specifications

Average Flow Rate Measurement

Technique ........................................................... Flow is calculated by measuring volume over time

Range ................................................................. 0.5 ml/h to 1000 ml/h

Accuracy ............................................................. 1 % of reading ±1 LSD for flows of 16 ml/h to 200 ml/h for volumes

over 20 ml; otherwise 2 % of reading ±1 LSD for volumes over 10 ml
under laboratory conditions

Max test duration ................................................ 10 hours on battery

Volume Measurement

Technique ........................................................... Volume is measured directly by the measuring module in minimum

sample sizes of 60 µl

Range ................................................................. 0.06 ml to 999 ml

Accuracy ............................................................. 1 % of reading ±1 LSD for flow rates of 16 ml/h to 200 ml/h for volumes

over 20 ml. Otherwise 2 % of reading ±1 LSD for volumes over 10 ml
under laboratory conditions

Max test duration ................................................ 10 hours on battery

Pressure Measurement

Technique (Occlusion test) ................................. Direct measurement of pressure at the inlet port

Range ................................................................. 0 psi to 45 psi and equivalent in mmHg, Bar and kPa

Accuracy ............................................................. 1 % of Full Scale ±1 LSD under laboratory conditions

Max test duration ................................................ 30 minutes