Infusion Device Analyzer

Product Maintenance

19

Clean the Product (Outside)

To clean the outside of the Product, disconnect from the power supply and use
only a damp cloth with mild detergent.

Clean the Product (Inside)

It is possible that microbial growth can become present in the measuring module
of the Product. It is recommended that you clean the fluid paths at 3 month
intervals. To clean the inside of the Product, inject 20 ml of a warm water and
detergent solution into the Fluid Inlet Port. After 5 minutes, flush with clean water.
Always pass water from the fluid inlets to the outlets.

Replaceable Parts

Table 11 lists the replaceable parts for the Product.

Table 11. Replaceable Parts

Item

Fluke Biomedical

Part Number

IDA-1S Infusion Device Analyzer

4468525

Hydrograph software and IDA-1S Users Manual on CD

4418071

Plastic syringe (20 ml)

4497350

3-way Luer plastic stop-cock

4480194

Drain tube (1 m)

4478942

Micro-90

 (225 ml)

4541948

USB Cable

1740487

Power supply, 4-plug mains adapter kit for US, UK, EURO, AUS

2461300

IDA-1S power supply cord without mains adapters, universal wall mount

4329971

NiMH Battery

4481150

IDA-1S Getting Started Manual

4426198